Add: Room 2001、2002、2003、2011, Zhonghai International Plaza No. 237, Wannianquan Road, Qingdao, Shandong 266199 China
Tel: +86 532 8281 2277
Mobile : +86 186 5320 8877(WhatsApp)
Email: market@hiprove.cn lena@hiprove.cn

Part parameters
- Packaging Details:
- Delivery Time:
- Description: MJ-560B2 Anesthesia Machine, designed with state-of-the-art technologies, is one of high utility with excellent performance and complete functions. Both outstanding and inexpensive, it has the best cost performance that can satisfy any customer.
Detailed introduction:
1.The Ventilator adopts microcomputer control with high-definition 5.7" LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
2.The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane,Isoflurane,
3. Sevoflurane, ensuring a successful anesthesia operation
4. Reliable pneumatic transmission system.
5. Economic and proper configuration of technologies, optional as required by the customer.
6. Modulatized design and upgrading. Upgrading is available as required by the customer.
| Anesthesia Machine | ||
| * | Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
| * | Scope of application | aduit, child |
| * | Gas source | O2: 0.27~0.55MPa N2O : 0.27~0.55MPa |
| * | Flow meter | O2:0.1~10L/min N2O : 0.1~10L/min |
| * | O2, N2O linkage and N2O Stopper | |
| When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. | ||
| * | Flow rate of rapid oxygen supply | 25~75L/min |
| * | Low oxygen pressure alam | There will be sound alarm when the oxygen pressure <0.2MPa |
| * | Vaporizer | |
| it has the function of automatic compensation based on pressure, temperature and flow rate, The regulation range of evaporator concentration is 0-5 vol%. Among Halothane, Enflurane, lsoflurane and Sevoflurane, one can be chosen for application as required by the customer.And vaporizers that are imported with original packaging are also available. | ||
| * | Respiratory circuit | Work mode: all-close, semi-close, semi-open release pressure: 1~ 4KPa |
| * | Respiratory Bellows | bellows for adults, bellows for children tidal volume range: 0~1500ml |
| Ventilator | ||
| * | Display mode | High-definition 5.7 " LCD screen display |
| * | Ventilation mode | |
| IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH | ||
| * | Ventilation parameters | |
| Tidal volume | 50~1500ml | |
| Rate | 2~99bpm | |
| SIMV rate | 2~20bpm | |
| I : E | 2:1~1:8 | |
| Inspiratory trigger pressure | -1.0~2.0KPa | |
| PEEP | 0~2.0KPa | |
| Pressure range | 1.0~6.0kpa | |
| SIGH | 1.5 times the inspiratory time | |
| * | Parameters for ventilation monitoring | |
| tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate,I/E, peak pressure of airway, pressure - time waveform ,PEEP, inspiratory trigger pressure |
||
| * | Monitoring of oxygen concentration | 21%~100% |
| * | Security Alert System | |
| Oxygen Concentration alarm | upper limit setting range 50%-100% | |
| low limit setting range 15%-50% | ||
| Airway pressure alarm | upper limit setting range 1.0-6.0kpa | |
| low limit setting range 0.4-2.0kpa | ||
| Per - minute ventilation volume alarm | upper limit setting range 3.0-30L/min | |
| low limit setting range 1.0-10L/min | ||
| Sustained high - pressure alarm | it will give alarm when stress have | |
| consistently been higher than 2.5 kpa | ||
| Suffocation alarmPower alarm | it will give sound and light alarm if there is | |
| no tidal volume input for 15 seconds | ||
| * | Power | AC 220v±10% 50Hz ±1HZ |
| ( UPS, Storage battery ) | ||
| * | Items within brackets are optional | |